Loading the rotor


Check the rotor for firm seating.

Remove the cover from the rotor:
Press the button (F) in the handle of the cover downward, and at the same time, pull the outer ring (G) of the
handle upward and remove the cover from the rotor. For figure, see the chapter "Installation and removal of the
rotor".

The rotor may only be loaded symmetrically. The centrifuge containers have to be distributed evenly on all rotor
positions. For authorised combinations see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and
accessories".
Example:
Rotor is evenly loaded

For the haematocrit measurement the capillaries are always filled up to  75%.

When loading the rotor, no liquid may enter the rotor or the centrifuging chamber.

When using standard capillary tubes, insert these in the rotor with the lute side out.
When using self-sealing capillary tubes, insert these in the rotor with the plug side out.

After loading the rotor, place the cover on the rotor and press downward on the handle until the cover engages.
AB1801DEENFRIT
Installation:

Place the rotor vertically onto the motor shaft (A) and press it
downwards. The motor shaft dog (B) has to fit in the rotor slot (C).
The alignment of the groove is labelled on the rotor (D).

Screw the fixing nut (E) with the supplied hex. pin driver on the motor
shaft (A) by turning in a clockwise direction and tighten it.
The fixing nut (E) must always be firmly tightened to the
motor shaft (A).

Check the rotor to make sure it is seated firmly.

Put the lid onto the rotor and press it down at the handle as far as the
C
lid clicks in.
A
Removal:

Press the knob (F) in the handle of the lid downwards,
simultaneously pull the outer ring (G) of the handle and remove the
B
lid from the rotor.

Unscrew the fixing nut (E) with the supplied hex. pin driver by turning
in a counter clockwise direction, and remove it.

Lift up the rotor from the motor shaft (A).
Rotor is not evenly loaded
Rev. 07 / 05.2020
Not permitted!
EN
49/134