5.4

Setting up the data connection (optional)

5.5

Aligning the scale

2 .
3 .
54 • seca 685
If you want to set up a data connection via
your hospital technician, your IT administrator or seca Service.
CAUTION!
Faulty measurement as a result of devices being set up incorrectly
Soft floors (wooden boards, for example) give under the patient's weight and fal-
sify the measuring result. Incorrect ambient conditions or lack of alignment lead
to measurement errors.
► Select a setup location with a level, stable floor in order to achieve precise
measuring results.
► Use the device only in the intended ambient conditions outlined in the section
entitled "Technical data".
► Align the device before using it and every time it changes location.
CAUTION!
Faulty measurement as a result of devices being set up incorrectly
If the weighing platform or attachments (e.g. ramps, handrails, rail parts) come
into contact with walls or objects, the measuring result will be falsified.
► Locate the device at a sufficient distance from walls and objects.
1. Place the scale on a firm, level surface.
2. Undo the knurled wheels.
4 .
3. Align the device by turning the foot screws.
The air bubble in the spirit level must be located in the exact center of the circle.
4. Tighten the knurled wheels in the direction of the arrow.
ð The foot screws are secured against being adjusted.
and
seca connect 103
seca 452
, contact