Activating Autohold
function (AHOld)
Activating/deactivating
beeps (bEEP)
Setting filtering (FIL)
When you activate the
Autohold
weighing procedure after the weight has been removed from the scale. It is then no
longer necessary to activate the
NOTE
This function is activated at the factory on some models. The function can be
deactivated.
1. From the menu, select the
2. Confirm your selection.
ð The current setting is displayed.
3. Select the desired setting:
–
On
–
OFF
4. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
You can activate beeps for a number of device functions:
1. From the menu, select the
2. Confirm your selection.
3. Select a menu item:
–
: beep when key is pressed
PrESS
–
HOLd
: beep when weighing value is stable
4. Confirm your selection.
ð The current setting is displayed.
5. Select the desired setting:
–
On
–
OFF
6. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
Setting the filter (
= filter) allows you to reduce interference when determining weight.
FIL
The selected setting affects the sensitivity with which the weight display reacts to pa-
tient movements and the time until the
nently (
value).
Hold
Setting (
FIL
)
Sensitivity
0
Sensitive
Medium
1
2
Insensitive
NOTE
With the
setting and patients who are not very steady on their feet, it is possi-
0
ble that no weight value is displayed permanently despite the
ing activated.
The greatest deviation between weight value displayed and actual value occurs
at setting
2
.
1. From the menu, select the
2. Confirm your selection.
ð The current setting is displayed.
3. Select a filtering level.
4. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
function, the measuring result is displayed for every
function manually.
Hold
menu item.
AHOLd
menu item.
bEEP
function displays a weight value perma-
Hold
Determination of
value
Slow
Medium
Fast
menu item.
FIL
English
Hold
function be-
Hold
Operation • 61