5.4
Setting up the data connection (optional)
5.5
Making the scale operational
Setting up the column
Aligning the scale
2 .
3 .
If you want to set up a data connection via
your hospital technician, your IT administrator or seca Service.
1. Place the scale on a solid, level surface.
2. Set up the column:
a) Undo the locking screw
b) Keep the locking screw pulled out
c) Set up the column
d) Engage the locking screw
e) Tighten the locking screw
CAUTION!
Faulty measurement as a result of devices being set up incorrectly
Soft floors (wooden boards, for example) give under the patient's weight and fal-
sify the measuring result. Incorrect ambient conditions or lack of alignment lead
to measurement errors.
► Select a setup location with a level, stable floor in order to achieve precise
measuring results.
► Use the device only in the intended ambient conditions outlined in the section
entitled "Technical data".
► Align the device before using it and every time it changes location.
CAUTION!
Faulty measurement as a result of devices being set up incorrectly
If the weighing platform or attachments (e.g. ramps, handrails, rail parts) come
into contact with walls or objects, the measuring result will be falsified.
► Locate the device at a sufficient distance from walls and objects.
1. Place the scale on a firm, level surface.
2. Undo the knurled wheels.