Error codes Description
Er 24 Heating plate temperature has ex- • SAFE TEMP H (Probe) is set a tem-
ceed the SAFE TEMP H (Hotplate) perature lower than the heating plate
setting
Er 25 Heating switching element
monitoring
Er 26 Difference between the internal • Irregular temperature distribution on Heating OFF
heating plate control and safety
temperature sensors is too large dissipation
Er 31 Fault in the heating switch
element (Triac)
Er 41 Stirring bar breakaway
WD 1 In remote operation (PC) with
(Watch watchdog function 2 enabled:
Dog)
No communication between PC
and RET control
Maintenance and cleaning
The device is maintenance-free. It is subject only to the natural
wear and tear of conponents and their statistical failure rate.
Cleaning
For cleaning disconnect the main plug!
Only use cleaning agents recommended by IKA
Contamination
Cleaning agents
Dyes
isopropyl alcohol
Building materials water containing tenside/isopropyl alcohol
Cosmetics
water containing tenside/isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
- When cleaning please do not allow moisture to get into the unit.
- Wear the proper protective gloves during cleaning of the devices.
- Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with IKA
that this method does not destroy the instrument.
Cause
actual temperature
• Breakage of the heating plate tempe-
rature control sensor
• There is a short-circuit in a switching Heating OFF
element (Triac) of the heating circuit
• Safety relay has shutdown the heating
circuit
• The heating or the connection is
broken
• The heating plate safety temperature
sensor is broken
the heating plate due to point heat
• Fault in the control or safety tempe-
rature sensors
• Stirring bar has broken away on four The setpoint speed is
consecutive occasions without any
changes to the setpoint value
• PC does not send OUT_WDx@m
data during the waatchdog time
• The connection/cable to the PC is
broken
.
®
®
Effect
Heating OFF
Heating OFF
reduced to 300 rpm
(min.)
Heating OFF
The setpoint temperature • Change the watchdo time
is set to the WD safety • Send data from the PC within the
temperature
The setpoint speed is set
to the WD safety speed
Spare parts order
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Software Version (second displayed value when the device is on)
- Item and designation of the spare part please see www.ika.com.
Repair
Please send in instrument for repair only after it has been cleaned
and is free from any materials which may constitute a health haz-
ard.
Please request the form to "clearance certificate" from IKA down-
load or use the printout of the form on the IKA
ika.com.
If you require servicing, return the instrument in its original pack-
aging. Storage packaging is not sufficient. Please also use suitable
transport packaging.
Solution
• Allow the heating plate to cool
• Set SAFE TEMP H (Hotplate) to a
higher temperature
• Switch off the unit. Allow it to cool
down and then switch on again
• Switching off the unit. Allow it to cool
down and then switch on again
• When using metal blocks on the heating
plate surface, ensure regular heat
dissipation
• Contact service
• Select a lower speed setpoint
• Choose smaller stirring bar
watchdog time (OUT_WD@m)
• Check the connector cable and plug
®
website: www.