• Do not ground the transducer, amplifier and indicator more than once. All
the devices in the measurement chain must be connected to the same
grounded conductor.
• The connection cable shielding is connected to the transducer housing.
• Connection diagram, grounding concept (Greenline).
Grounding concept (Greenline)
The cable shield is connected in accordance with the Greenline concept. This
encloses the measurement system in a Faraday cage. Any electromagnetic
interference active here does not affect the measurement signal.
In the case of interference due to differences in potential (compensating cur
rents), the connection between operating voltage zero and the housing ground
must be separated at the amplifier and a potential equalization line estab
lished between the housing and the amplifier housing (highly flexible stranded
2
wire, 10 mm
wire cross‐section).
6
Maintenance
The TB2 reference torque transducers are maintenance‐free.
7
Option
• Degree of protection IP67 per EN 60529
8
Accessories
To be ordered separately:
• Connection plug MS 3106 PEMV, mounted on cable
• 15‐pin D‐plug, mounted on cable
• In combination with DKD calibration certificate, Class 0.05 as per DIN