Herunterladen Diese Seite drucken

HBM TB2 Montageanleitung Seite 16

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

16
5
Electrical connection
The torque reference transducer is supplied with a ready‐made 6‐wire trans­
ducer connection cable with free ends. A plug can be attached on request
(see Chap. LEERER MERKER).
Extension cables should be shielded and low capacitance. HBM provides spe­
cific cables for this purpose, the 1‐KAB0304A‐10 (ready‐made) and the
KAB8/00‐2/2/2 (by the meter, can also be supplied with mounted device con­
necting plugs).
The pin assignment can be found in the following table:
Connection
Measurement signal (+U
Bridge excitation voltage (-U
Bridge excitation voltage (+U
Measurement signal (-U
Sense lead (-)
Sense lead (+)
Shielding connected to enclo­
sure ground
Table 5.1 Pin assignment
For pin assignments of amplifiers with solder or clamped connections, please
refer to the documentation of the applicable amplifier.
Fig. 5.1: PIN assignment Lemo
HBM
PIN
Wire color
)
6
wh (white)
A
)
1
bk (black)
B
)
5
bu (blue)
B
)
3
rd (red)
A
2
gy (gray)
4
gn (green)
1
6
5
2
3
4
R
plug
Connecting to an amplifier with
15‐pin
SUB‐D connector
8
5
6
15
12
13
Top view (from outside)
TB2
Plug MS3106
PEMV
A
B
C
D
G
F
A0884-5.0 en/de

Werbung

loading