8. The 16EZA holder can accommodate from one to four BV800R capsules or one BV5600R
capsule. If more than one BV800R capsule is needed, place the second capsule on top of the
fi rst capsule with the female connector facing up.
9. Repeat step 8 until all the capsules are loaded.
10. Place the top manifold on top of the last capsule that is loaded. The male connector of the top
manifold will face downward, the upstream and downstream vents will face forward, and the
manifold stops will be positioned between the support rods when the top manifold is in the
correct orientation.
11. Turn the hand wheel clockwise until you feel the torque limiter slip and moveable end plate is
stationary.
4.B. Production-Scale Holder (16EZB) Loading Instructions for Single Stage Purifi cation
Loading is performed while facing the front of the empty holder with the capsule compression
wheel on your right, with the holder in the horizontal position.
1. Ensure holder is on a level surface and engage both fl oor locks.
IMPORTANT NOTICE: Holder must be in the horizontal position for loading.
2. Pull the locking pin out and rotate the locking bar to the unlocked position.
3. Slide the left end plate to the far left.
IMPORTANT NOTICE: When installing seven (7) capsules, the locking bar must be touching
the rubber bumpers
4. Standing facing the right end of the holder, fully retract the right end plate by rotating the
capsule compression wheel counterclockwise.
5. Standing to the front of the holder place the bottom manifold against the right end plate with the
female connector facing left, sanitary connectors facing upward and manifold stops positioned
between notched rods.
6. Lubricate all o-rings using water or process fl uid.
IMPORTANT NOTICE: Inspect the o-rings and carry out a hardware qualifi cation prior to use.
7. Place the capsule in the holder so that the capsule stops rest against the front notched rod.
8. Slide the capsule to the right until it touches the bottom manifold.
9. Push the capsule handles away from you until the straight handle stops against the back
notched rod.
10. If using more than one capsule, place the next capsule in the holder so that the capsule stops
rest on the front notched rod.
11. Slide the capsule to the right until it touches the previously installed capsule.
12. Push the subsequent capsule handles away from you until the straight handle stops against the
back notched rod.
13. Repeat steps 10 - 12 until all the required capsule fi lters are installed.
14. Place the top manifold next to the last capsule installed with the male connector facing right,
sanitary connectors facing upward and manifold stops positioned between notched rods.
15. Standing to either the front or left side of the holder, slide the left end plate until the locking bar
is positioned over the notch closest to the top manifold and lock.
16. Rotate both manifolds to the desired locations within notched rods.
17. Standing facing the right end of the holder, turn the capsule compression wheel clockwise until it
comes to a complete stop (a minimum torque of 100 inch-lbs [11.3 Nm] is required) and there are
no visible gaps between all capsules and manifolds.
18. Standing to the back of the holder, turn the rotation wheel clockwise to pivot the holder to the
vertical operating position.