Herunterladen Diese Seite drucken

Storz UI 500 ENDOFLATOR 50 Gebrauchsanweisung Seite 88

Werbung

80
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Table 3 – Test levels for Radiated and conducted Immunity Tests
Guidance and manufacturer's declaration – electromagnetic immunity
The ENDOFLATOR
50 is intended for use in the electromagnetic environment specified below. The user of ENDOFLATOR
®
is used in such an environment.
Immunity tests
EN/IEC 60601 test level
Conducted RF disturbances
3 V
rms
acc. to IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF disturbances acc.
3 V/m
to IEC 61000-4-3
80 MHz to 2.5 GHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of buildings,
objects, and persons.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the ENDOFLATOR
applicable RF compliance level above, the ENDOFLATOR
measures may be necessary, such as re-orienting or relocating the ENDOFLATOR
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Electromagnetic
Compatibility (EMC)
information
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the ENDOFLATOR
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
f
6 V
rms
where P is the nominal power of the transmitter in watts [W] accor-
ding to the information provided by the manufacturer of the transmit-
ter and d is the recommended separation distance in meters [m].
Field strengths from fixed RF transmitters, as determined by an elec-
tromagnetic site survey
each frequency range
f
3 V/m
h
Interference may occur in the vicinity of equipment marked with the
following symbol:
7
50 should be observed to verify normal operation. If abnormal performance is observed, additional
®
50.
®
Indicaciones sobre
compatibilidad
electromagnética (CEM)
50 should make sure that it
®
50, including cables, than
®
a
should be less than the compliance level in
.
b
80 MHz to 800 MHz
800 MHz to 2.5 GHz
50 is used exceeds the
®

Werbung

loading

Diese Anleitung auch für:

Endoflator 50Ui 500