Samoa / Samoa lite
4. Service and Maintaining the device
4.1. Charging the battery
Caution: It is not allowed to use a different
battery charger than the one provided!
The charger must not be operated in the patient's
environment!
4.1.1. Charger
A power supply unit is available for charging the
battery. This is specially matched to the device's
battery.
Battery charging procedure :
Connect the battery charger to the device
•
socket marked with the symbol
Plug the charger into a power outlet
•
Device with display: The integrated LED's in
•
the top cover of the device displays the char-
ging status of the batteries.
Device without display: The charging condi-
tion of the battery is displayed at the top of
the device
The device can remain connected to the charger ad-
aptor for a long period of time without suffering any
damage.
4.1.2. Charging Cradle
To charge the battery optionally, the battery
compartment on the back of the device can be
opened and the battery removed. The battery
charging adaptor is specially designed for this type of
battery and must only be used with the device
battery. It is recommended to charge the battery
completely after each measurement.
Battery charging procedure :
Remove the battery from the device and insert it
•
into the battery cradle.
Plug the charger into a power outlet
•
The charger LED will illuminate orange. As soon
•
as the battery is charged, the LED will illuminate
green. The battery can remain in the charging
cradle for a longer period of time without dam-
age.
4.2. Cleaning instructions
General
As with any medical device, when using the Samoa
system, certain hygienic procedures are required for
safe reuse on the patient. Reusable products must
have safe disinfectability to exclude any risk of
infection from subsequent users / patients. The
regulations of the Medical Devices Act stipulate that
for such medical devices disinfection activities with
procedures according to the RKI guideline are to be
carried out in analogy to the methods of surface or
instrument disinfection. Sterilization of the products
listed below is not necessary.
The Samoa system is a medical device that is in the
patient's environment when used as intended, with
direct contact with the hands of the staff and the
patient. These products are therefore classified as
so-called
"non-critical
accordance with the guidelines for hospital hygiene
and infection prevention.
The cleaning and disinfection instruction for the
device can be found in the "Hygiene Plan Sonata /
Scala / Samoa", which can be requested from
authorized dealers.
Warning:
No liquid should enter the device or plug connec-
tions. The sensors must not be immersed in liquid.
Before cleaning, close the pressure connections of
the device with caps!
Disconnect the charging and data cable from the
device before each cleaning.
In the washing machine
The thorax and abdominal straps can be washed in
the washing machine at 60 °C (use a laundry net!).
Let the straps dry in the air.
Attention:
Do not use a dryer
•
Remove thoracic and abdominal sensor
•
beforehand (not Samoa lite)
Disposables
Disposables must not be used multiple times or on
different patients. Follow the guidelines for waste
disposal in hospitals.
4.3. Maintenance
The device is maintenance-free. It is recommended
to have a technical inspection carried out by the
manufacturer or an agency authorized by the
manufacturer every 24 months.
In order to maintain basic safety and essential
performance, no special measures need to be taken
with regard to EMC.
4.4. Transport and storage conditions
The ambient temperatures for transport and storage
are between -25 °C and +70 °C, where:
-25 °C and +5 °C without restrictions,
•
+ 5 °C to +35 °C with a relative humidity of
•
up to 90%, without condensation,
> 35° C to 70 °C at a water vapor pressure
•
of up to 50 hPa.
Condensation must always be avoided.
5. Evaluation software for PC
5.1. Installing the software
To install the PC software, place the supplied CD into
the CD-ROM drive or insert the supplied flash drive
into the USB port of your computer. The program will
automatically run. If the program does not run
automatically open the Explorer on your operating
system and select the CD-ROM drive or flash drive.
- 25 EN -