Modell 2700 Resource Block – Referenz
Tabelle C-1.
Resource Block Parameter Fortsetzung
Parameter-Mnemonik
78
RECOMMENDED_ACT
ION
216
Meldungs
Definition
typ
Spezifizierte Liste empfoh-
VARIABLE
lener Maßnahmen,
angezeigt mit einem
Gerätealarm.
Datentyp/
Grö-
Struktur
ße
unsigned16
2
D
-
Auswerteelektronik Modell 2700 mit F
Liste der Werte
R
0 = Uninitialized
1 = No action
6 = Factory configuration
checksum invalid
7 = Factory configuration
invalid
8 = Electronics Failure -
Device
9 = Replace the Fieldbus
Electronics Module
Assembly
10 = Transmitter
Initializing/Warming Up
11 = Reset the Device
then Download the Device
Configuration
12 = Sensor
Communication Failure
13 = Low Power
14 = No Sensor Response
15 = Mass Flow
Overrange
16 = Density Overrange
17 = No Left Pickoff/Right
Pickoff Signal
18 = Temperature
Overrange
19 = CM: Unable to Fit
Curve Data
20 = Transmitter Not
Characterized
21 = Calibration Failure
23 = Data Loss Possible
(Totals)
24 = Drive Overrange
25 = PM: Temperature or
Density Overrange
26 = Extrapolation Alert
27 = Meter Verification
Failed
28 = Meter Verification
Aborted
29 = Slug Flow
30 = Sensor Simulation
Active
32 = Allow the procedure
to complete
33 = Check Transducer
Block Mode
34 = Simulation Active
39 = Simulated -- Factory
configuration checksum
invalid
40 = Simulated -- Factory
configuration invalid
41 = Simulated --
Electronics Failure -
Device
42 = Simulated -- Replace
the Fieldbus Electronics
Module Assembly
43 = Simulated --
Transmitter
Initializing/Warming Up
44 = Simulated -- Reset
the Device then Download
the Device Configuration
45 = Simulated -- Sensor
Communication Failure
46 = Simulated -- Low
Power
47 = Simulated -- No
Sensor Response
48 = Simulated -- Mass
Flow Overrange
49 = Simulated -- Density
Overrange+
50 = Simulated -- No Left
Pickoff/Right Pickoff
Signal
51 = Simulated --
Temperature Overrange