Manufacturer's Declaration

WARNING: The use of accessories, transducers, and cables other than those
specified may result in increased emission and/or decreased immunity of ARGUS
VCM monitor.
ARGUS VCM
customer and/or user of
electromagnetic environment as described below;
Harmonic emissions
Voltage fluctuations/flicker
Emission Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
IEC 61000-4-11
is suitable for use in the specified electromagnetic environment. The
ARGUS VCM
Table 12. Electromagnetic Emissions (IEC60601-1-2)
Emission Test
Compliance
RF emission
Group 1
CISPR 11
RF emissions
CISPR 11
Complies
IEC 61000-3-2
Complies
emission
IEC 61000-3-3
Table 13. Electromagnetic Immunity (IEC60601-1-2)
IEC60601-1-2 Test
Level
±6 kV contact
±8 kV air
±2 kV for
power supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV
common mode
<5 % U T
(>95 % dip in UT )
for 0.5 cycle
40 % U T
(60 % dip in UT )
for 5 cycles
70 % U T
(30 % dip in UT )
for 25 cycles
ARGUS VCM
should assure that it is used in an
Electromagnetic Environment
must emit electromagnetic
ARGUS VCM
energy in order to perform its intended
function. Nearby electronic equipment may
be affected.
Class B
is suitable for use in all
ARGUS VCM
establishments.
Compliance
Electromagnetic Environment
Level
±6 kV contact
Floor should be wood,
±8 kV air
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
±2 kV for
Mains power quality should be
power supply
that of a typical commercial
lines
and/or hospital environment
±1 kV for
input/output lines
±1 kV differential
Mains power quality should be
mode
that of a typical commercial
±2 kV
and/or hospital environment
common mode
<5 % U T
Mains power quality should be
(>95 % dip in U T)
that of a typical commercial
for 0.5 cycle
and/or hospital environment. If
the user of
requires continued operation
during power mains
interruption, it is
40 % U T
recommended that
(60 % dip in U T)
for 5 cycles
VCM
uninterruptible power supply
70 % U T
or battery.
(30 % dip in UT)
for 25 cycles
Operator's Manual
Specifications
Guidance
ARGUS VCM
ARGUS
be powered from an