Troubleshooting
Serviceable error codes and other error codes are listed in
the monitor continues to present an error code, call the Schiller technical representative and
report the error code number. You will be advised of the remedial action to be taken. Before
calling the Schiller Technical Services Department, make sure that the battery is charged,
and that all power connections are correctly made.

Corrective Action

If you experience a problem while using
qualified service personnel or Schiller Technical Services Department.
ARGUS VCM
additional troubleshooting information.
Following is a list of possible errors and suggestions for correcting them.
1. There is no response to the Power On/Off switch.
A fuse may be blown. Notify service personnel to check and, if necessary, replace
the fuse.
If operating on battery power, the battery may be missing or discharged. If the
battery is discharged, charge the battery, see Battery Operation section.
2. The monitor display does not function properly and the power-on beep tones do not
sound during the power-on self-test.
Do not use
Services Department.
3. The monitor is operating on battery power, even though it is connected to AC.
Make sure that the power cord is properly connected to
Check to see if power is available to other equipment on the same AC circuit.
The monitor will be operated from its internal battery if in doubt about the integrity
of the AC power source,
84
ARGUS VCM
service manual, which is for user by qualified service personnel, provides
; contact qualified service personnel or Schiller Technical
ARGUS VCM
Operator's Manual
ARGUS VCM
and are unable to correct it, contact
ARGUS VCM
ARGUS VCM
ARGUS VCM
service manual. If
.