Specifications
Emission Test
Power
frequency
(50/ 60 Hz)
magnetic field
IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.
Emission Test
Potable and mobile RF communications equipment should be used no closer to any part of
ARGUS VCM , including cables, than the recommended separation distance calculated from
the equation appropriate for the frequency of the transmitter.
Radiated RF
IEC 6100-4-3
Conducted RF
IEC 61000-4-6
Note: Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast,
and TV broadcast cannot be predicted theoretically with survey accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which ARGUS VCM is
used exceeds the applicable RF compliance level above, ARGUS VCM should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Interference may occur in the vicinity of equipment marked with the following symbol:
94
ARGUS VCM
IEC60601-1-2 Test
Level
<5 % U T
(95 % dip in UT )
for 5 sec.
3 A/m

Table 14. Electromagnetic Immunity (IEC60601-1-2)

IEC60601-1-2
Test Level
3 V/m, 80 MHz
3 V/m
800 MHz
3 V/m
3 V/m
800 MHz
2.5 GHz
3 Vrms
3 Vrms
150 kHz
80 MHz
Operator's Manual
Compliance
Electromagnetic Environment
Level
<5 % U T
(95 % dip in UT)
for 5 sec.
3 A/m
It may be necessary to position
ARGUS VCM
sources of power frequency
magnetic fields or to install
magnetic shielding. The power
frequency magnetic field
should be measured in the
intended installation location
to assure that it is sufficiently
low.
Compliance Level
Electromagnetic Environment
Recommend Separation Distance
distance = 1.2 √Power
distance = 2.3 √Power
80 MHz to 2.5 GHz
distance = 1.2 √Power
150 kHz to 80 MHz
.
ARGUS VCM
ARGUS VCM
Guidance
further from the
Guidance