Measurement Parameters - Schiller Argus VCM Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Specifications

Measurement Parameters

NIBP
Pulse Rate Range
Pulse Rate Accuracy
Technique
Measurement modes
AUTO Mode
MANUAL Mode
STAT Mode
NIBP pressure measurement range
Systolic pressure range
Diastolic pressure range
Mean pressure range
Pressure Display Accuracy
Cuff Pressure Range
Initial Cuff Inflation
Overpressure protector
Standards
Note: Systolic and diastolic blood pressure measurements determined with this device are equivalent to
those obtained by a trained observer using the cuff/stethoscope auscultation method, within the limits
prescribed by the American National Standard, Electronic or automated sphygmomanometers.
90
ARGUS VCM
Pulse Rate
Adult/Pediatric/Neonatal 30 BPM to 220 BPM
±3 BPM or ±3%, whichever is greater
NIBP (Non-Invasive Blood Pressure)
Oscillometric Measurement
AUTO, MANUAL and STAT
Automatic NIBP measurements at intervals of 1, 2, 3, 4, 5, 10, 15,
30, 45, 60, 90, 120, and 240 minutes
Single measurement initiated by NIBP Start/Stop switch
Series of consecutive measurements for 5 minutes
Adult:
Pediatric: 30 mmHg to 160 mmHg (3.9 kPa to 21.2 kPa)
Neonatal: 25 mmHg to 120 mmHg (3.3 kPa to 15.9 kPa)
Adult:
Pediatric: 15 mmHg to 130 mmHg (1.9 kPa to 17.2 kPa)
Neonatal: 10 mmHg to 105 mmHg (1.3 kPa to 13.9 kPa)
Adult:
Pediatric: 15 mmHg to 140 mmHg (1.9 kPa to 18.6 kPa)
Neonatal: 10 mmHg to 110 mmHg (1.3 kPa to 14.6 kPa)
Meets ANSI/AAMI SP10:1992+A1:1996
0 to 300 mmHg (0 to 40 kPa)
Adult:
100, 120, 140, 160(Default), 180, 200, 220, 240, 270 mmHg
(13.3, 15.9, 18.6, 21.2(default), 23.9, 26.6, 29.2, 31.9, 35.9 kPa)
Pediatric:
80, 90, 100, 110, 120(default), 130, 140, 150, 160, 170 mmHg
(10.6, 11.9, 13.3, 14.6, 15.9(default), 17.2, 18.6, 19.9, 21.2, 22.6 kPa)
Neonatal:
50, 60, 70, 80, 90(default), 100, 110, 120, 132 mmHg
(6.6, 7.9, 9.3, 10.6, 11.9(default), 13.3, 14.6, 15.9, 17.5 kPa)
Adult 280 mmHg (N.C.), 308 mmHg (S.F.C.)
Pediatric 220 mmHg Neonatal 154 mmHg
ANSI/AAMI SP10:1992+A1:1996, IEC60601-2-30:1999
EN1060-1:1995 and EN1060-3:1997.
Operator's Manual
30 mmHg to 260 mmHg (3.9 kPa to 34.5 kPa)
20 mmHg to 235 mmHg (2.6 kPa to 31.2 kPa)
20 mmHg to 255 mmHg (2.6 kPa to 33.9 kPa)
ARGUS VCM