Setting NIBP Units

This procedure will allow you to select a NIBP measurement unit either mmHg or kPa.
With the monitor in the normal monitoring mode:
1. Press and hold the Mode button for 3 seconds or more until the monitor enters
2. Select a NIBP unit either mmHg or kPa, using the Up/Down (+/-) selection
3. Press any other button to return to normal operation. If there is no activity for
3 seconds, the monitor will return to normal operation.
28
ARGUS VCM
Configuration mode. Once the monitor is in the Configuration mode, current
NIBP unit is flashing on the display.
buttons.
Figure 11. NIBP Units Setting