RS-232 Interface

Nurse Call Interface

WARNING: Although the Nurse Call option enables remote notification of an alarm
condition, it is not intended to replace appropriate bedside patient surveillance by
trained clinicians.
The monitor can be connected to a Nurse Call system through a customized cable that
connects to the Nurse Call connector. When the cable is connected and operational, the
monitor immediately notifies the Nurse Call system when a patient alarm occurs.
The Nurse Call function specification is as follows:
72
ARGUS VCM
Switch current:
1 A maximum
Switch voltage:
30 Vac/dc maximum
Isolation:
1500 Vrms
Alarm relay:
Energized during patient alarm
Operator's Manual