Alarms and Limits

Setting Alarm Limits

During patient monitoring, an alarm occurs when a measurement falls outside the
programmed alarm limit. Alarms can be set or turned off for the following vital signs:
The range of high and low alarm limits for each vital sign is shown here:
Parameter
Systolic (mmHg)
Neonatal
Pediatric
Adult
Diastolic (mmHg)
Neonatal
Pediatric
Adult
MAP (mmHg)
Neonatal
Pediatric
Adult
PR (bpm)
Neonatal
Pediatric
Adult
SpO2 %
Neonatal
Pediatric
Adult
56
ARGUS VCM
Systolic high and Systolic low alarm limits
Diastolic high and Diastolic low alarm limits
MAP high and MAP low alarm limits
Pulse rate high and Pulse rate low alarm limits
SpO 2 high and SpO 2 low alarm limits
Table 8. Range of Alarm Limits (Defaults)
Low Limit, Default
25 to 115, 50
30 to 155, 75
30 to 255, 75
10 to 100, 30
15 to 125, 35
20 to 250, 35
10 to 105, 35
15 to 135, 50
20 to 250, 50
25 to 295, 100
25 to 295, 50
25 to 295, 50
50 to 98, 85
50 to 98, 90
50 to 98, 90
Operator's Manual
High Limit, Default
Resolution
30 to 120, 100
5 mmHg (0.7 kPa)
35 to 160, 145
5 mmHg (0.7 kPa)
35 to 260, 220
5 mmHg (0.7 kPa)
15 to 105, 70
5 mmHg (0.7 kPa)
20 to 130, 100
5 mmHg (0.7 kPa)
25 to 255, 110
5 mmHg (0.7 kPa)
15 to 110, 80
5 mmHg (0.7 kPa)
20 to 140, 110
5 mmHg (0.7 kPa)
25 to 255, 120
5 mmHg (0.7 kPa)
30 to 300, 200
5 bpm
30 to 300, 150
5 bpm
30 to 300, 120
5 bpm
52 to 100, 98
1%
52 to 100, 100
1%
52 to 100, 100
1%
ARGUS VCM