Following procedure will allow you to reset the monitor operating parameters to the
factory default settings.
With the monitor powered off:
1. Simultaneously press the Power button and the NIBP start/stop button.
2. The monitor runs a self-test and then displays the current monitor software
version.
3. Press the Mode button 4 times until " DEFAULT RESET=NO" is displayed.
4. To leave the settings unchanged, select 'NO' using only the Down (-) selection
button.
To reset the operating parameters to the factory default values, select 'YES'
using only the Up (+) selection button. The monitor immediately resets to the
defaults and the confirmation tone sounds.
5. To return to normal operation, turn the monitor off; then restart.