Alarms and Limits
Diastolic high and low alarm limits
With the monitor in the normal monitoring mode:
1. Press the Alarm button 3 times until Diastolic high alarm limit is displayed.
2. Leave the limit unchanged or press the Up/Down (+/-) selection buttons as needed
to change the limit to another value.
3. Press the Alarm button once again until Diastolic low alarm limit is displayed.
4. Leave the limit unchanged or press the Up/Down (+/-) selection buttons as needed
to change the limit to another value.
58
ARGUS VCM

Figure 32. Diastolic High Alarm Limit Setting

Figure 33. Diastolic Low Alarm Limit Setting
Operator's Manual