Aesculap Power Systems
microspeed® uni motor system

2. Product description

2.1 System components

Designation
microspeed® uni control unit with cooling unit
Bottle holder
Instructions for use
Notes on Electromagnetic Compatibility (EMC)

2.2 Components necessary for using the product

• microspeed® uni control unit with cooling unit GD670
• Power cord
• Motor
• Motor cable
(when using a motor with a motor-side plug connection)
• Foot control
(when using a motor cable without hand control, or the shaver
handpiece)
• Bottle holder
(when using the coolant pump)
• Tubing set
(when using the coolant pump)
• Surgical suction device
(when using the shaver handpiece)

2.3 Intended use

The microspeed® uni motor system is a universal system with respect to
its range of applications and to the system composition and its variety of
uses.
The microspeed® uni motor system is used in orthopedics/traumatology,
neurosurgery, spine surgery, hand/foot surgery, ENT/OMF surgery, plastic
surgery and arthroscopy.
The microspeed® uni motor system can be operated as complete from the
sterile area, either through foot control or through hand control.
4

2.4 Mode of operation

The microspeed® uni motor system comprises the following components:
• Control unit with coolant pump, for universal application
• Two motor cables (with or without hand control)
• Various motors with different coupling types
• Pistol handpiece
Art. no.
• Shaver handpiece
• Two foot control devices
GD670
GD412804

2.4.1 Control unit

TA022089
Control unit 1 has been designed for a mains voltage rating of 100 V to
TA022130
240 V/50 Hz to 60 Hz. The mains voltage is transformed so that the motors
run at safety low voltage.
Control unit 1 is equipped with two motor connection sockets 5 for
connecting two (different) motors, and with one foot control connection
socket 14 (on rear panel).
The following devices can be connected to control unit 1 at the same time:
• Two motors and
• One foot control (GD668 or GD671).
Note
Do not connect two shaver handpieces GD682 at the same time.
Only one motor can be activated at any one time.
Motors with motor-side plug connection (GD674, GD676, GD677, GD678
and GD679) can be operated either through hand control (motor cable
GD673) or foot control GD668 or GD671 (with motor cable GD672).
Note
When two motors without hand control are connected to the control unit
(motor cable GD672), a motor can only be activated if the motor release
switch of the motor that is to be activated is in the "ON" position while the
other motor is locked (motor release switch in "OFF" position).

2.4.2 Control and display interface

Display 2 shows the present device status (operating and error conditions)
at all times. The display is divided into motor control field(s) and a pump
control field. The device parameter settings can be changed by pressing
the control keys.
For information on all the symbols used in the motor and pump control
fields, see chapter Symbols in display.
Control key "Instruction menu" calls up the Instruction menu.