CLM280
12.3 Measuring
12.3.1 Overview of the measuring modes
12.3.2 Special functions
AutoRead
(drift control)
AutoRange
(measuring range selection)
Temperature measure-
ment
Reference temperature The reference temperature (Tref) can be switched between 20 °C
Endress+Hauser
AutoRead checks the stability of the measurement signal.
1. The function is enabled with <AR>.
The current measured value is frozen (Hold function).
2. Press <RUN/ENTER> to start the AutoRead measurement.
During the measurement, AR flashes on the display until a
stable measured value is reached.
3. You can abort the measurement with <RUN/ENTER>.
The current measured value is accepted.
If a measuring range is exceeded, AutoRange causes the measuring
instrument to change automatically to the next measuring range.
The function is permanently switched on (display shows ARng).
The fluid temperature is automatically measured by the integrated
temperature sensor of the conductivity sensor (display shows TP).
The measuring instrument automatically recognizes the type of
temperature sensor used. You can connect sensors with NTC30 or
Pt1000.
(68 °F) and 25 °C (77 °F). It appears on the display as Tref20 or
Tref25.
Switching the reference temperature --> CONFIGURING.