CLM280
Measured variable
Measuring range
Reference operating
conditions
Maximum measured
2)
error
Temperature
compensation
Cell constant of the
sensor supplied
2)
In accordance with DIN IEC 746 Part 1, at reference operating conditions
Endress+Hauser
16
Technical data
16.1 Input
Conductivity [S/cm] or resistivity [MΩ
• Conductivity:
0.0 to 199.9 μS/cm | 0 to 1999 μS/cm | 0.00 to 19.99 S/cm |
0.0 to 199.9 mS/cm | 0 to 500 mS/cm
• Resistivity:
0.00 to 19.99 MΩ
• Salinity
0.0 to 70.0 as per IOT (International Oceanographic Tables)
• Temperature:
-5 to 100 °C (20 to 212 °F)
16.2 Performance characteristics
Reference temperature:
Reference pressure:
Conductivity:
Resistivity:
Salinity:
Temperature:
Nonlinear temperature compensation as per EN 27 888 with adjustable
reference temperature
-1
0.475 cm
.
cm]
.
cm | 0.0 to 199.9 MΩ
25 °C (77 °F)
1013 hPa (15 psi)
0.5 % of the measured value ± 1
digit
0.5 % of the measured value ± 1
digit
0.2 % ± 1 digit
0.1 K ± 1 digit
Technical data
.
.
cm | 0 to 1999 MΩ
cm
29