1.
Press <CAL> until the following appears:
2.
Press <RUN/ENTER>.
3.
Press <CAL> to switch the temperature compensation on and off (nLF or - - - -).
4.
Switch back to the measuring mode with <M>.
The following table is intended as a reference point for selecting temperature compensation de-
pending on the type of medium involved:
Medium
Natural water
Other aqueous solutions
Sea water
1)
International Oceanographic Tables

12.6 Configuring

!
Note!
You can leave the configuration menu at any time with <M> . The parameters that have already
been changed are stored.
1.
Switch off the measuring instrument.
2.
Press <M> and <
3.
Press <CAL> to select the reference temperature for calculating the conductivity.
You can choose from:
– Tref 20 (reference temperature 20 °C (68 °F))
Endress+Hauser
Type of compensation
DIN 38404 / EN 27888
None
Automatically nLF accord-
1)
ing to IOT
I
> simultaneously.
Display
nLF
- - - -
Sal, nLF
Commissioning