4.4 Notes on digital Setup

1
"Single" = single range, "Dual I" = dual interval, "Dual r" = dual range and
"Direct" = direct mV/V calibration without weights.
2
Sets the decimal position before weight range input.
3
Enter the full scale weight (Max) for single interval, or (Max1) for dual interval/range.
Use units of displayed weight (not divisions).
4
Enter the verification interval (e) for single interval, or (e1) for dual interval/range. Use
units of displayed weight.
5
Enter (Max2) for dual interval/range.
6
Enter e2 for dual interval/range.
7
Additive Tare limit in weight. Up to this weight the WE2110 operates as a additive Ta-
ring system. ! The max. load of the scale (e.g. CAP1) is exceeded by that value !
8
The weight unit set here will be used for the display and any printer output.
9
Trade use has some limitations. See description of commands for that.
10
This is the number of readings that will be averaged for display.
11
Anti-jitter prevents minor display instability with small average filter settings (see 10).
12
Motion detection rate is set in the number of e1 per 0.5 second or per second.
13
Enables the initial zero setting at power-on. The fixed setting range is −5% ... +15%.
14
Zero Tracking rate is set in the number of e1 per 0.5 second or per second.
15
This sets the range of operation of the ZERO setting.
16
The band around zero is defined. This is used by the zero track function
and the printout.
17
Deactivate the error message E0040...E00C0. Useful for EX-I applications.
18/19
See page 62 for separate chart for Zero and Span Calibration. Some digital settings
can effect calibration, always complete digital Setup before calibrating. Enter zero and
span in mV/V with direct-mV/V-calibration.
20/21
Linearity Routine. See diagram page 63 (not possible with mV/V adjustment).
22
Cancels existing calibration and restores original factory default calibration.
23
The Serial 1 output can be set to OFF, Auto.Lo (approx. 10/sec), Auto.Hi (ca. 50/sec),
Single transmit by remote key, or Network.
24
The Serial 2 output can be set for OFF, Auto.Lo (approx. 10/sec), Single transmit, or
Printer driving.
25
The network address can be set from 00 to 31.
26
Type of automatic output can be set. Auto.A is the standard HBM format, Auto.B, Au-
to.C or Auto.D are extended with units fields etc. See the description of commands for
more information about automatic output formats.