Herunterladen Inhalt Inhalt Diese Seite drucken

Delivery Checklist; Initial Operation; Opening And Closing The Lid; Opening The Lid - Hettich EBA 270 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für EBA 270:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7

Delivery checklist

The following items and accessories are delivered with the centrifuge:
1 Connecting cable
2 Fuses
1 Release pin
1 One-opend-end wrench
1 Hex. pin driver for transport protection
1 Rotor
6 Hanger 10 ml
6 Hanger 5 ml
1 Notes on moving the equipment safely
1 Operating instructions
8

Initial operation

Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Substances can be discharged through the ventilation opening on the rear side of the
centrifuge. The device is to be placed in such a way that the air stream is not directed towards
people.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specifications".
Turn on the mains switch. Switch position "Ι".
The last used centrifuge data will be displayed.
Open the lid.
Remove the transportation safety device in the centrifuge, see sheet "Transportation safety device".
9

Opening and closing the lid

9.1

Opening the lid

The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
Press the key
and open the lid. The symbol " " (lid open) illuminates in the rotation indicator
OPEN
9.2

Closing the lid

Do not bang the lid shut.
Place the lid and lightly press down the front edge of the lid. The symbol " " (lid closed) illuminates in the rotation
indicator
.
AB2300DEENFRIT
Rev. 11 / 12.2022
EN
.
37/85

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis