Cleaning and disinfecting the facepiece
1. Remove any accessories, such as mask spectacles, and
clean them separately if required.
2. Prepare a cleaning solution consisting of water and a
cleaning agent.
3. Clean all parts using a soft cloth with cleaning solution.
4. Thoroughly rinse all parts under running water.
5. Prepare a disinfectant bath consisting of water and a
disinfectant.
6. Put all parts to be disinfected in the disinfectant bath.
7. Thoroughly rinse all parts under running water.
8. Dry all parts in the air or in a drying cabinet (temperature:
max. +60 °C). Do not expose to direct sunlight.
9. Fit the accessories.
The facepieces of the X-plore 5500 and X-plore 6000
series can also be cleaned and disinfected using a machine.
Further information can be obtained from Dräger.
3.4

Leak test

General notes
Carry out the leak test with a suitable test unit (e.g. from the
Testor or Quaestor series).
Moistening the face seal facilitates exact donning, since the
mask can slide onto the test head more easily.

3.4.1 Checking for leaks

1. Moisten the face seal with water.
2. Moisten the exhalation valve with water.
3. Mount the facepiece on the test head.
4. Make sure that the face seal fits tightly onto all parts of the
test head. If necessary, inflate the test head.
5. Seal the front port (X-plore 6000 series) or bayonet-type
connector (X-plore 5500) with the adapter and produce a
negative pressure of 10 mbar.
The facepiece is considered leak-tight if the pressure drop
is less than 1 mbar/min.
If the facepiece is leaking, consult the following chapter
(see "Test with sealed exhalation valve", page 9).

3.4.2 Test with sealed exhalation valve

1. Remove the exhalation valve disc.
2. Insert the exhalation valve test plug.
3. Produce a negative pressure of 10 mbar.
4. If the pressure drop is less than 1 mbar/min, remove the
exhalation valve test plug and insert a new valve disc.
5. Repeat the leak test (see "Checking for leaks", page 9).

3.4.3 After testing

1. Disconnect the adapter.
2. Remove the facepiece from the test head and dry it if
necessary.
3. Put on the exhalation valve cover.
The protective cap must lock into place.
Technical manual
|
Dräger X-plore 5500Dräger X-plore 6000
3.5

Replacing the speech diaphragm

1. Disconnect the inner mask.
2. Unscrew the speech diaphragm with the speech
membrane wrench and remove it.
3. Visually inspect the speech diaphragm and O-ring and
replace them if necessary.
4. Insert the O-ring, then screw in the speech diaphragm.
5. Attach the inner mask.
3.6

Replacing the exhalation valve

1. Open the exhalation valve cover.
2. Disconnect the valve disc.
3. The valve disc and valve seat must be clean and
undamaged. Otherwise clean or replace them.
4. Fit the valve disc.
5. Close the exhalation valve cover.
3.7

Replacing the inhalation valve

3.7.1 X-plore 5500

1. Disconnect the inner mask.
2. Pull the interior inhalation valve from the valve seat.
3. The valve discs and valve seat must be clean and
undamaged. Otherwise clean or replace them.
4. Connect the valve to the valve seat.
5. Reattach the inner mask.

3.7.2 X-plore 6000 series

1. Pull the inhalation valve seat with valve out of the front
port. Do not damage the front port.
2. The valve disc and valve seat must be clean and
undamaged. Otherwise clean or replace them.
3. Insert the inhalation valve seat with valve into the
facepiece.
|
en
Maintenance work
9