Maintenance
Following any maintenance work and/or replacement of
components, repeat the leak test.
4.1
Recommended test and maintenance
intervals
The following are recommendations from the DGUV-R 112-190, which
applies in Germany. Observe national directives.
Type of work to be performed
Check by the device user
Visual inspection, function and
leak test

Cleaning and disinfecting

Replace the exhalation valve disc
Replace speech diaphragm
1)
Every 2 years for full face masks in airtight packaging.
2)
For masks in airtight packaging; otherwise every 6 months.
18
X
1)
X
X
2)
X
X
X
X
(approved by Dekra Exam)
4.2
Cleaning and disinfecting
Clean and disinfect the full face mask after every use.
WARNING
!
The mask body must be cleaned sufficiently thoroughly that
no residues of skin care products remain on the mask body.
Otherwise the elasticity in the sealing area can be lost and the
full face mask will no longer sit correctly.
CAUTION
!
Do not use solvents (e.g. acetone, alcohol) or cleaning agents
containing abrasive particles for cleaning and disinfection.
Only clean the mask using the described methods and
cleaning/disinfecting agents. Other agents, dosages and
contact times can damage the product.
Do not clean or dry the inside of full face masks with visors
marked "PC/CC" with a cloth; this may damage the coating.
Full face masks with these visor types should be left
unpackaged after drying for at least a day in a normal climate
according to DIN EN ISO 291 so that the coating can
regenerate and the optimum service life can be achieved.
For information about suitable cleaning agents and
disinfectants and their specifications, refer to document
9100081 at www.draeger.com/IFU.
1. If necessary, remove accessories and clean them separately.
2. Prepare a cleaning solution consisting of water and a cleaning
agent.
3. Clean all parts using a soft cloth with cleaning solution.
4. Thoroughly rinse all parts under running water.
 Prepare a disinfectant bath consisting of water and a disinfectant.
5. Put all parts to be disinfected in the disinfectant bath.
6. Thoroughly rinse all parts under running water.
Panorama Nova ESA