Immunity Test
Magnetic field at the
supply frequency
(50/60 Hz)
acc. to IEC 61000-4-8
Radiated,
radio-frequency,
electromagnetic field
immunity test
acc. to IEC 61000-4-3
Immunity to conducted
disturbances, induced
by radio-frequency
fields
acc. to IEC 61000-4-6
IEC/EN 60601 test level
30 A/m at 50Hz/60Hz
3V/m 80 MHz to 2.7 GHz
*Refer table 2 for
wireless Proximity RF
field test levels
3Vrms on 150 kHz to
80MHz 1kHz 80% AM
modulation
6Vrms in ISM bands
Compliance level
30 A/m at 50Hz/60Hz
3V/m 80 MHz to 2.7 GHz
3Vrms on 150 kHz to
80MHz 1kHz 80% AM
modulation
6Vrms in ISM bands
9
Electromagnetic environment
- guidance
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
Environment.
Refer to Table 3
Refer to Table 3