mobile RF communications equipment and the Clinician Programmer
The Clinician Programmer is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or user of the Clinician Programmer can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Clinician Programmer as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance D
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
NOTE: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected
by absorptions and reflections of structures, objects and people.
Table 4 – Recommended separation distances
Recommended separation distances between portable and
Separation distance d [m] as a function of the transmitter frequency
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
12
800 MHz to 2.7 GHz
d = 2.3 √ P
0.23
0.73
2.3
7.3
23