Guidelines and manufacturer's declaration – electromagnetic emissions
The Clinician Programmer is intended for use in the electromagnetic environment specified below.
The customer or user of the Clinician Programmer should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF Emissions as per CISPR 11
Harmonic emissions as per
IEC 61000-3-2
Voltage fluctuations/flicker
Emissions as per IEC 61000-3-3
Table 6: Test levels for RFID Immunity testing to AIM 7351731
Frequency
134.2 KHz
13.56 MHz
433.92 MHz
860-960 MHz
2.45 GHz
Table 5 – Emission class and group
Compliance
Group 1
Class A
Class A
Complies
Immunity Test Level
65 A/m
5 A/m - 12 A/m
3 V/m
54V/m
54 V/m
Electromagnetic environment - guidance
The Clinician Programmer uses RF energy only for its
internal function. The customer or user of the Clinician
Programmer should make sure that it is used in such
an environment. Therefore, its RF emissions are very
low and are not likely to cause interference in nearby
electronic equipment.
The Clinician Programmer is suitable for use in all
establishments other than domestic and those directly
connected to the public low voltage power supply
network that supplies buildings used for domestic
purposes.
13
Compliance Level
65 A/m
5 A/m - 12 A/m
3 V/m
54V/m
54 V/m