Maintenance

9.1

Replacing the control unit fuse

Users can replace faulty control unit fuses themselves. These are located at the rear of the device in the fuse slot
beside the power switch:
Unplug the power plug.
• Open the fuse slot using a screw driver.
• Replace defective fuses T 1 AL 250 VAC.
• Slide the fuse holder back in and close the fuse slot.
• Plug in the power plug again.
1
1.
Fuse slot locking mechanism
2. Fuse slot
3. Fuse 1
4. Fuse 2
9.2

Safety inspections

The performance of safety inspections on medical devices is required by law in several countries. The safety inspec-
tion is a regular safety check that is compulsory for those operating medical devices. The objective of this measure is
to ensure that device defects and risks to patients, users or third parties are identified in time.
An inspection interval of 2 years applies to the Vacuson 60 LP.
NOUVAG AG offers a safety inspection service for its customers. Addresses can be found in the appendix of this
operation manual under "Service centers". For further information please contact our technical service department.
Further international service centers are listed on the Nouvag website:
9.3

Bacteria filter

A periodical replacement of the bacteria filter is recommended after 8 hours of use, but definitely after it came into
contact with foam or infectious material. For reordering refer to chapter 11 to retrieve the article number.
16
2
www.nouvag.com > Service > Service centers
Operating instructions Vacuson 60 LP, Art. No. 31793, V46/13