Herunterladen Inhalt Inhalt Diese Seite drucken

Replacement Of The Measuring Circuit Fuse; Rear Panel; Scanner Card Ho112 - Hameg HM8112-3 Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für HM8112-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

26
A/SOURCE (4mm safety socket)
Connection of measuring cables for:
current measurement, max. 1 ampere
2 wire resistance measurement
4 wire resistance measurement (SOURCE)
4 wire temperature measurement by a PT-temperature
sensor (SOURCE)
continuity test up to 10 Ω
The maximum current may amount to 1 A
LOW (4mm safety connectors)
25
Ground connection for inputs
high-impedance DC-isolated.
27
Fuse in the current measuring circuit
The shunt is fuse-protected. The fuse (FF) is located in a fuse
holder. The measuring circuit is designed for a maximum al-
lowable measurement current of 1 ampere.
Replacement of the fuse is only allowed, after the
instrument was disconnected from the mains!
A repair of a defective fuse or bypassing the fuse is
very dangerous and absolutely prohibited!

10.9 Replacement of the measuring circuit fuse

The measuring circuit fuse
27
A replacement of the fuse is only allowed, if no voltage is applied
to the measuring connectors! Therefore all terminals V/SENSE
24
, ground
25
and A/SOURCE
cover of the fuse holder has to be turned ccw with a screw driver
having a suitable blade. As the cover can be turned it has to be
pushed by the srew driver into the fuse holder. The cover with
the fuse can then be easily taken out. Replace the defective fuse
by a new fuse of the same type having the same trip current. A
repair of a damaged fuse or the use of other means for bypassing
the fuse is very dangerous and absolutely prohibited! Damages
incurred will void the warranty.

10.10 Rear Panel

28
Power receptacle with power switch
Power receptacle for connecting the line cord with according
to DIN49457.
29
Interface
The USB/RS-232 interface is located at the rear panel of the
HM8112-3. The interface of HM8112-3 can receive data (com-
mands) from an external device (PC) or send data (measurement
values and parameters). The following option is available: HO880
IEEE-488 (GPIB). In order to avoid the warranty seal broken we
recommend the installation ex factory.
30
Voltage selector
Choice of mains voltage (115 V / 230 V).
29
!
eff
24
and
26
. Both connectors are
is accessible from the front panel.
26
should be disconnected. The
28
30
S c a n n e r C a r d H O 1 1 2
11 Scanner Card HO112 (option)
Miscellaneous:
With built-in Scanner Card HO112 voltage measurements are only
possible up to 100 V. That means that the 600 V range of the voltage
measurement functions is automatically inactive.
Pin 1 is the ground connection. Channel BP is used to supply the
other channels with current, e.g. for suppling sensors, LEDs etc.
Commands:
03A0
all channels are off
03A1
channel 1 active
03A2
channel 2 active
03A3
channel 3 active
03A4
channel 4 active
03A5
channel 5 active
03A6
channel 6 active
03A7
channel 7 active
03A8
channel 8 active
03A9
front channel
active
Pin assignment:
AHI
VHI
AHI
VHI
AHI
VHI
AHI
VHI
19
18
17
16
15
14
13
12
11
37
36
35
34
33
32
31
30
ALO
VLO
ALO
VLO
ALO
VLO
ALO
VLO
ALO
BP
CH1
CH2
CH3
Specifications
Channels:
8 (4-wire)
Switching:
bistable, floating relais
Thermal voltage:
typ. 500 nV, max. 1µV*)
Max. voltage between 2 contacts: 125 V
Max. measuring voltage:
125 V
Volt-Hertz-Product:
≤ 1 x 10
Max. switching current:
1 A
Max.contact resistance:
approx. 1 Ω (each wire)
Life time:
2 x 10
Insulating resistance:
3 GΩ **)
Capacity:
>100 pF, between contacts
Switching delay:
20 ms
Measurement delay:
between 50 ms and 300 ms
*) max. 1µV after a warm-up of 1.5 h
**) at rel. humidity < 60 %
AHI
VHI
AHI
VHI
AHI
VHI
AHI
VHI
AHI
10
9
8
7
6
5
4
3
29
28
27
26
25
24
23
22
21
VLO
ALO
VLO
ALO
VLO
ALO
VLO
ALO
CH4
CH5
CH6
CH7
CH8
pk
- also V/Ω-input -
pk
6
V · Hz
eff
8
switches (0.1 A; 10 V
)
DC
Subject to change without notice
VHI
2
1
20
VLO
59

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hm 8112-3

Inhaltsverzeichnis