6½-Digit Precision Multimeter
HM8112-3 [HM8112-3S]
All data valid at 23 °C after 30 minutes warm-up.
DC specifications
Ranges HM8112-3:
0.1 V; 1 V; 10 V; 100 V; 600 V
Ranges HM8112-3S:
0.1 V; 1 V; 10 V; 100 V
Input impedance:
0.1 V, 1.0 V
>1 GΩ
10 V, 100 V, 600 V
10 MΩ
Accuracy Values given are in ±(% of reading (rdg.) + % of full scale (f.s.)):
1 year;
23 °C ±2 °C
Range
% rdg.
0.1 V
0.005
1.0 V
0.003
10.0 V
0.003
100.0 V
0.003
600.0 V
0.004
Integration time:
0.1 s
Display range:
120.000 digit
600 V range
60.000 digit
Resolution:
1 µV
Zero point:
Temperature drift
better than 0.3 µV/°C
Long-term stability
better than 3 µV for 90 days
AC specifications
Ranges HM8112-3:
0.1 V; 1 V; 10 V; 100 V; 600 V
Ranges HM8112-3S:
0.1 V; 1 V; 10 V; 100 V
Measurement method:
true rms, DC or AC coupled
(not in 0.1 V range)
Input impedance:
0.1 V, 1 V
1 GΩ II <60 pF
10...600 V
10 MΩ II <60 pF
Response time:
1.5 sec to within 0.1 % of reading
Accuracy:
For sine wave signals >5 % of full scale
Values given are in ±(% of reading + % of full scale); 23 °C ±2 °C for 1 year
Range 20 Hz...1 kHz 1...10 kHz
0.1 V
0.1+0.08
5+0.5 (5 kHz)
1.0 V
0.08+0.08
0.15+0.08
10.0 V
0.08+0.08
0.1+0.08
100.0 V
0.08+0.08
0.1+0.08
600.0 V
0.08+0.08
0.1+0.08
Temperature coefficient 10...21 °C and 25...40 °C; (% rdg. + % f.s.)
at 20 Hz...10 kHz
0.01 + 0.008
at 10...100 kHz
0.08 + 0.01
Crest factor:
7:1 (max. 5x range)
Integration time:
0.1s
Display range:
120.000 digit
600 V range
600.00 digit
Resolution:
1 µV
Overload protection (V/Ω-HI to V/Ω-LO) and to chassis:
Measurement ranges:
all
all the time:
850 V
Maximum input voltage LOW against chassis/safety earth:
250 V
Current specifications
Ranges:
100 µA; 1 mA; 10 mA; 100 mA; 1 A
Integration time:
0.1 s
Display ranges:
120.000 digit
1 A range
100.000 digit
Resolution:
1 nA
Accuracy:
DC
(1 year; 23 °C ±2 °C)
0.02 + 0.002
Temperature coefficient/°C:
10...21 °C
(%rdg. + %f.s.)
0.002+ 0.001

Specifications

Voltage:
<600 mV...1.5 V
Response time:
1.5 s to within 0.1 % of reading
Crest factor:
7:1 (max. 5 x range)
Input protection:
fuse, FF 1 A 250 V
Resistance
Ranges:
100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ
Integration time:
0.1 s
Display ranges:
120.000 digit
Resolution:
1 mΩ
Temp. coefficient
% f.s.
10...21 °C + 25...40 °C
0.0006
0.0008
0.0006
0.0008
0.0006
0.0008
0.0006
0.0008
0.0006
0.0008
1...60 s
1,200.000 digit
600.000 digit
100 nV
10...50 kHz 50...100 kHz 100...300 kHz
0.3+0.1
0.8+0.15
7+0.15
0.3+0.1
0.8+0.15
4+0.15
0.3+0.1
0.8+0.15
1...60s
1,200.000 digit
600.000 digit
100 nV
or 600 V
peak
dc
at max. 60 Hz or 250 V
rms
dc
1...60 s
1,200.000 digit
1,000.000 digit
100pA
45 Hz...1 kHz
1...5 kHz
0.1 + 0.08
0.2 + 0.08
25...40 °C
0.01+ 0.01
1...60 s
1,200.000 digit
100 µΩ
Accuracy Values given are in ±(% of reading + % of full scale):
1 year;
23 °C ±2 °C
Range
%rdg
100 Ω
0.005
0.0015
1 kΩ
0.005
0.001
10 kΩ
0.005
0.001
100 kΩ
0.005
0.001
1 MΩ
0.05
0.002
10 MΩ
0.5
0.02
Measurement current:
Range
100 Ω, 1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
Max. measurement voltage: approx. 3 V
Overload protection:
250 V
Temperature measurement
PT100/PT1,000 (EN60751):
2- and 4-wire measurement
Range
-200...+800 °C
Resolution
0.01 °C; measurement current 1 mA
Accuracy
±(0.05 °C + sensor tolerance + 0.08 K)
Temperature coefficient
10...21 °C and 25...40 °C
<0.0018 °C/°C
NiCr-Ni (K-type):
Range
-270...+1,372 °C
Resolution
0.1 °C
Accuracy
±(0.7 % rdg. + 0.3 K)
NiCr-Ni (J-type):
Range
-210...+1,200 °C
Resolution
0.1 °C
Accuracy
±(0.7 % rdg. + 0.3 K)
Frequency and period specifications
Range:
1 Hz...100 kHz
Resolution:
0.00001...1 Hz
Accuracy:
0.05 % of reading
Measurement time:
1...2 s
Specification Scanner Card HO112: refer to page 59
Interface
Interface:
Dual-Interface USB/RS-232 (HO820),
IEEE-488 (GPIB) (optional)
Functions:
Control / Data fetch
Inputs:
Function, range, integration time, start
command
Outputs:
Measurement results, function, range,
integration time (10 ms...60 s)
Miscellaneous
Time to change range or
approx. 125 ms with DC voltage, DC current,
function:
resistance approx. 1 s with AC voltage,
AC current
Memory:
30,000 readings /128 kB
Safety class:
Safety class I (EN 61010-1)
Power supply:
105...254 V~; 50...60 Hz, CAT II
Power consumption:
approx. 8 W
Operating temperature:
+5...+40 °C
Storage temperature:
-20...+70 °C
Rel. humidity:
5...80 % (non condensing)
Dimensions (W x H x D):
285 x 75 x 365 mm
Weight:
approx. 3 kg
*)
max. 1 µV after a warm-up of 1.5 h
**)
at rel. humidity <60 %
Accessories supplied: Line cord, Operating manual, PVC test lead (HZ15),
Interface cable (HZ14), CD
Recommended accessories:
HO112
Scanner Card (Installation only ex factory) as HM8112-3S
HO880
Interface IEEE-488 (GPIB), galvanically isolated
HZ10S
5 x silicone test lead (measurement connection in black)
HZ10R
5 x silicone test lead (measurement connection in red)
HZ10B
5 x silicone test lead (measurement connection in blue)
HZ13
Interface cable (USB) 1.8 m
HZ33
Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34
Test cable 50 Ω, BNC/BNC, 1.0 m
HZ42
19" Rackmount kit 2RU
HZ72
GPIB-Cable 2 m
HZ887
Temperature probe
S p e c i f i c a t i o n s
Temp. coefficient/°C
%f.s.
10...21 °C
25...40 °C
0.0008
0.0008
0.0008
0.0008
0.0008
0.0008
0.0008
0.0008
0.002
0.002
0.01
0.01
Current
1 mA
100 µA
10 µA
1 µA
100 nA
p