8 Concept of the HM8112-3

8.1

Reference

The integrated AD converter has to be connected to a reference.
The characteristics of this reference determine the long term
stability of the instrument. The reference of HM8112-3 is there-
fore a high precision reference device.
8.2

Integrated AD converters

Converters applying the multi slope method are used for AD
conversion.
8.3

Moving average

The value determined by the AD converter could be displayed
without prior computations, also the average calculated from
n – values could be shown. First of all 1 to n values will be
logged. Averaging over these values will be done, and sub-
sequently this average will be displayed. After 120 values the
next value n+1 will be quantified by the AD converter. The pri-
mary measured value 1 will be abolished and a new average
will be calculated from the remaining values ( 2 to n) and from
the new value n+1. This has the advantage that peaks and in-
terferences will be smoothed.
8.4

Measurement of alternating values

Frequency, period
Frequency and period are both measured by a pulse-counting
circuit. Time base is 1 second. The first falling edge triggers
the measurement and starts the counter. For one second
every falling edge will trigger a counting pulse. After expiry of
this term the measurement circuit will wait for the next zero
point. Hence the signal's period is measured. The time will be
measured until the next zero point occurs. The measurement
result determines the frequency of the signal and the period will
I n t r o d u c t i o n t o t h e o p e r a t i o n o f t h e H M 8 1 1 2 - 3
be calculated from the frequency. This combined measurement
of the number of zero points and of the period of a signal allows
the measurement of very small as well as very high frequencies
within a reasonable time. Applying of a DC voltage results in a
frequency displayed of 0 Hz.
As the period is calculated from the measured frequency divi-
sion by zero will be made. Therefore the instrument will display
„INF" if the period of a DC voltage is measured („INF" = infinity).
RMS rectifier
The AC voltage is measured by a high precision RMS rectifier
device. This device gauges a DC voltage proportional to the
applied AC voltage. This DC voltage is equivalent to the true
RMS value of the AC voltage.
Measurement of the crest factor
For crest factors exceeding 7 an AC voltage or current measu-
rement will be incorrect due to the true RMS converter.

9 Introduction to the operation of the HM8112-3

Especially before the first operation please pay attention to the
following points:
The line voltage, indicated on the rear panel of the instru-
ment must correspond to the line voltage used.
Operation is only allowed from 3 terminal connectors with
a safety ground connection or via isolation transformers of
class 2.
No visible damage to the instrument.
No damage to the line power cord.
No loose parts in the instrument.

Factory settings

The following values are set by default:
The measurement range is
The sampling rate amounts to
The function „1:Filter" is
The temperature is displayed in
The selected temperature sensor is
The data logger is
The RS-232 interface is
10 V
DC
100 ms
OFF
°C
PT100
OFF
OFF
51