Carefully remove the insertion instrument A from
the implant a. Avoid any tilting or canting of the
instrument when doing so.
Place bone or bone replacement material around
the implant a.
Removing the implant intraoperatively
If, in the opinion of the surgeon, it becomes necessary
to remove the implant intraoperatively, carry out the
following steps:
Check the position for coupling the insertion
instrument A, see Fig. 1.
Couple insertion instrument A to implant a in situ,
see Fig. 7.
Turn the switch 2 to the right (
Fig. 3.
Mount the impact cylinder B onto the insertion
instrument A, see Fig. 5.
Carefully expel the implant a using the impact
cylinder C out of the intervertebral space, see
Fig. 6.
8
XP
instruments
Disassembling
Insertion instrument SN305R
direction), see
Fig. 12 Insertion instrument disassembled
Assembling
Turn the rotary nut 3 counterclockwise ("loosen"
direction).
Turn the switch 2 to the left (
Fig. 1.
Turn the rotary nut 3 clockwise ("fix" direction),
until the rod 6 is out of the thread.
Remove the rod 6 from the tube 1, see Fig. 12.
1
6
Turn the switch 2 to the left ( direction).
Insert the rod 6 into the tube 1, see Fig. 12.
Turn the rotary nut 3 counterclockwise ("loosen"
direction) to push the rod 6 forward, see Fig. 1.
direction), see