The following table includes possible failures and corresponding corrective measures.
Malfunction
On/off button illumi-
nation remains off
No mixing function
No heating function
Temperature display
not plausible with
temperature sensor
connected
Deviation setpoint/
current value of
sample
If a fault cannot be rectified with the described remedies, please contact an authorized
sales representative or our technical service (see section „Contact information Heidolph
international", page 46).
Possible cause/ remedy
Mains voltage not present: Check the power supply
cable for damage, check the connection plug for correct
seating, check the fuse of the house installation.
LED defective, contact technical service.
No magnetic stirring bar in the vessel, insert stirring
bars.
Rotation speed set to zero, increase rotation speed.
Heating temperature setpoint below current temperature,
adjust setpoint.
Heating plate defective, contact technical service!
Check temperature sensor connection.
The immersion depth is too low, note the minimum
immersion depth of 20 mm.
Setpoint too low, take into account heat losses.
Heating plate defective, contact technical service!