Operation
Heating function (Hei-PLATE Core only)
The following sections are only relevant for Hei-PLATE Core devices. Devices of the type
Hei-PLATE Mix 20 l do not offer heating function!
Setting the heating temperature
WARNING
Risk of burn
The heating plate is designed for a maximum operating tempera-
ture of 300°C. Contact with surfaces above 50 °C can lead to
serious injury.
➜
➜
➜
➜
Turn on the device as described in section „Switching the device on/off", page 37.
[temperature]
➜
Turn the
temperature (setting range: 20 – 300 °C).
Switching the heating function on/off
➜
Make sure the device is switched on and the heating temperature is set.
[heating on/off]
➜
Press the
–
When the heating function is activated, the LED ring of the
off]
button and the corresponding
➜
The heating temperature can be adjusted at any time during operation by turning
[temperature]
the
[heating on/off]
➜
Press the
Avoid direct skin contact with the heating plate during operation
and also observe the residual heat indicator after switching off!
Do not place heat-sensitive objects on the heating plate.
Always use the necessary personal protective equipment
(heat-resistant gloves, eye protection, safety clothing) to
process samples from a temperature of 50 °C.
control clockwise or counter-clockwise to set the heating
button to activate the heating function.
control.
again to turn off the heating function.
[heating]
indicator LED will light orange.
40
[heating on/