Herunterladen Inhalt Inhalt Diese Seite drucken

Setting The Rotation Speed - Heidolph Hei-PLATE Mix 20 l Betriebsanleitung

Vorschau ausblenden Andere Handbücher für Hei-PLATE Mix 20 l:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation

Setting the rotation speed

WARNING
Risk of injury, risk of damage to property caused by splash-
ing fluids
When using open vessels, there is a risk that fluid will spray out.
Turn on the device as described in section „Switching the device on/off", page 37.
[rotation]
Turn the
speed (setting range: 100 – 1,400 rpm).
The rotation speed can be adjusted at any time.
Whenever possible, use closed vessels for processing corrosive,
toxic or biohazardous substances and seal them safely.
Especially using open vessels, increase the rotation speed only
gradually and observe the fluid movements.
Hei-PLATE Core: Always use the necessary personal
protective equipment (heat-resistant gloves, eye protection,
safety clothing) to process samples from a temperature of
50 °C.
control clockwise or counter-clockwise to set the desired
39