6. Commissioning, operation
EN
1. Seal the sealing faces.
2. Screw the precision digital pressure gauge into the mounting point by hand.
When screwing in, do not cross the threads.
3. Tighten with a torque spanner using the spanner flats.
The maximum torque of the CPG1500 is 13.5 Nm = 10 ftlbs.
6.2

Electrical mounting

DANGER!
Danger to life from explosion!
Through working in flammable atmospheres, there is a risk of explosion
which can cause death.
6.3

Normal operation

Make a long press on the [On/Off] button, in order to activate the precision digital
pressure gauge. Push the button again to turn it off.
After switching on, the start-up screen with pressure range and firmware version is shown
on the display for approx. 3 seconds.
Zeroing the display: Press and hold the [ZERO] button.
The CPG1500 has to be set to 0 with the [ZERO] button before each use.
MAX/MIN: The CPG1500 stores the minimum and maximum pressure in the memory.
This value can be activated under "MENU / MEASURING MODE / PEAK VALUES" and
is displayed in the text field.
20
If the message "OL" is displayed, the measuring range has been
exceeded and the pressure source must be removed immediately from the
CPG1500 to prevent damage to the internal sensor.
Only carry out set-up work in non-hazardous environments!
Only connect test and calibration installations once the system has been
depressurised (atmospheric).
Do not use rechargeable batteries!
Only use the listed batteries, see chapter 3.6.2 "Permitted batteries"!
Only replace the batteries outside the hazardous area, see chapter
8.2 "Battery replacement"!
Always replace all three batteries together!
WIKA operating instructions, model CPG1500