Herunterladen Inhalt Inhalt Diese Seite drucken

Electromagnetic Compatibility Tables - I-Tech MIO-IONOTENS Benutzerhandbuch

Elektrotherapie
Vorschau ausblenden Andere Handbücher für MIO-IONOTENS:
Inhaltsverzeichnis

Werbung

ELECTROMAGNETIC COMPATIBILITY TABLES

Guidance and manufacturer's declaration – ELECTROMAGNETIC EMISSIONS –
MIO-IONOTENS is intended for use in the electromagnetic environment specified
below. The customer or the user of MIO-IONOTENS should assure that it is used in
such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonics emissions
IEC 61000-3-2
Voltage
fluctuation/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY –
MIO-IONOTENS is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IACER Srl
FOR ALL EQUIPMENT AND SYSTEMS
Compliance
Group 1
Class B
Class A
Compliant
FOR ALL EQUIPMENT AND SYSTEMS
Test level
Compliance level
IEC 60601
±8kV in contact
±8kV in contact
±2Kv, ±4kV,
±2Kv, ±4kV,
±8kV; +15kV on
±8kV; +15kV on
air
air
±2kV
±2kV
for power
for power
supplies lines
supplies lines
47
Electromagnetic environment -
guidance
MIO-IONOTENS uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
MIO-IONOTENS is suitable for
domestic establishment and in
establishment directly connected to
the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Electromagnetic
environment - guide
Floors
wood, concrete or
ceramic tile. If floors
are
synthetic
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
should
be
covered
with
material,
MNPG176-07