Gelenkarm
Inhaltsverzeichnis

Werbung

4. With one hand, pinch the sliding
u-piece until the imaging head
is firmly held in place.
5. With your other hand, turn the
knob to tighten and lock.
To loosen: Reverse the above steps.
9. Inspect Before Use
All products should be inspected throughout their lifetime to
ensure proper function and performance. If product does not
pass the inspection, do not use.
Main Arm (HH1001)
1. Visually inspect component for damage or cracks.
2. Check to make sure that the arm is rigid at all three joints by
turning central tightening knob clockwise.
3. Insert arm column into table clamp, turn column clamping knob
clockwise and ensure that it holds securely.
4. Check quick connect distal end to make sure the collar locks
and unlocks accessory securely.
Table Clamp (HH1002)
1. Visually inspect component for damage or cracks.
2. L-Bracket on Table Clamp should clamp securely to table rail.
3. Insert Main Arm into column opening, turn central tightening
knob clockwise and make sure the post is held securely inside
the column opening.
Imaging Head Holder (HH1003)
1. Visually inspect instruments for damage or cracks.
2. Quick connect shaft should freely engage and disengage in
and out of the quick connect distal end
©2020 IFU-0100 / rev3
9