6. Commissioning, operation
4. Using the CONFIG key search for the options mentioned above and lock and unlock
functions using the ▲ and ▼ keys. To scroll through the parameters press the CONFIG
key, and press the Enter key to save the settings and switch the pressure gauge back to
GB
normal operation. If a function is locked, it can neither be accessed nor can you change
the current status. To modify a locked function, enter the Supervisor Mode and unlock
the function. Once it has been modified, if necessary, enter the Supervisor Mode to lock
it again.
5. Setting of a user-defined unit or scale: The last menu option in the Supervisor Mode is
SET FACTOR. Using this option you can define a multiplier between 0.001 and 100 for
a user-defined unit. The defined factor is multiplied by the psi measured. The result is
displayed. Example: 40 psi corresponds to 1000 g of a product in a tank. The weight of
the product should appear on the display, a pressure gauge for 100 psi is used. If 25 is
defined as the multiplier, the pressure of 40 psi corresponds to the value 1000 (40 x 25).
"CUST" is displayed on the instrument as the engineering unit.

6.3 Normal operation

Turning on and off the background illumination: Press the Enter key.
Zeroing the display: Press and hold the Zero key.
MAX/MIN: The CPG1000 stores the minimum and maximum pressure in the memory. If
the MAX/MIN key is pressed once, the maximum pressure from the memory is displayed.
If the MAX/MIN key is pressed again, the minimum pressure from the memory is displayed.
After 2 seconds the instrument returns to normal operation. To delete the memory registers
holding the values for MAX/MIN, hold the MAX/MIN key pressed for at least 2 seconds
until the "CLr" message is displayed. The analogue progress bar display on the lower
edge of the display indicates the present pressure proportional to the entire measuring
range. Remember that, if a Tare value has been programmed for the pressure gauge, the
displayed pressure does not correspond to the actual pressure present.
22
WIKA Operating Instructions Model CPG1000