Error Messages

Messages

Errors

Fatal Errors

Message
Process in Progress.
Process already stopping.
Water level in tank too low; or pump
filter clogged.
Water level low; emptying finished.
Flush nozzle not parked.
Flushing was interrupted to avoid
pump damage. Please park the
flush nozzle to continue.
Emptying of tank was stopped to
avoid pump damage.
Valve handle in emptying tank pos.
Do not operate handle now.
Reference position not found.
Cutting stopped by flange sensor.
Exotom-100
Instruction Manual
Error messages are divided into three classes:
Messages
Errors
Fatal Errors
Messages are intended to inform the operator of the machine's
progress and advise about minor operational errors.
Errors must be rectified before cutting can be continued.
In case of Fatal Errors, cutting cannot continue before an
authorised technician has rectified the error. Turn off the unit at
the Main Switch immediately. Do not attempt to operate unit
before technician has rectified problem.
Explanation
Start is not possible as other
process in progress e.g. emptying
of recirculation tank.
Appears if Stop is pressed
repeatedly.
Water level in recirculation tank is
too low. Alternatively, recirculation
jumper filter clogged.
Emptying of recirculation cooling
unit completed.
Start not possible. Start was
pressed while flush nozzle was out
of holder.
The pump automatically stops after
3 minutes to avoid pump damage.
The recirculation pump
automatically stops after 8 minutes
to avoid pump damage.
Drain pump handle has not been
moved to off position after use.
During cutting, an attempt is made
to activate tank emptying function
manually.
Stop was pressed during search for
reference point.
Flange sensors next to the cut-off
wheel flange will stop the
movement of the cutting arm if it
meets an obstacle.
Action
Refill circulation water if needed.
Check condition of pump filter.
Clean if needed.
Switch off recirculation pump to
complete emptying process.
Park flush nozzle to continue.
Park flush nozzle to continue.
To re-start the pump, turn the pump
handle horizontally, then re-activate
the pump by turning the pump
handle to the vertical position.
Re-position drain pump handle.
Press Start again to do new search.
If message remains, switch unit off
and on at Main Switch to start new
search for reference point.
Replace cut-off wheel or re-position
workpiece.