Message
Disc motor overloaded
Sample motor overloaded
Pump motor overloaded
Main supply voltage too low
Removal rate too low, phase time
exceeded
Sample holder unable to move,
upwards after process
Force regulation error
Air pressure too low
Stock removal sensor not installed
58
AbraPol-10
Instruction Manual
Explanation
The disc motor was overloaded and
therefore overheated.
The sample motor was overloaded
and therefore overheated.
Water hose blocked or too much
swarf in coolant.
Supply voltage insufficient.
It has not been possible to
grind/polish the samples to the
expected level within 15 minutes.
It was not possible to move the
sample holder upwards after
finishing a process.
It is not possible to reach the
selected force.
Missing or incomplete air supply.
Stock removal sensor has not been
installed or has not been calibrated.
Action
Wait till the motor has cooled off
and reduce the force before
restarting the process.
Wait till the motor has cooled off
and reduce the force before
restarting the process.
Replace coolant. Inspect pump.
Await normalisation of power grid.
Select a different surface or/and
increase the force.
The pressure in the compressed air
system is too low.
Force regulation error – contact
Struers service technician.
The pressure in the compressed air
system is too low.
Force regulation error – contact
Struers service technician.
Check that the air hoses are tight
and properly clamped.
If stock removal sensor (accessory)
is not installed this must be done
before the stock removal mode can
be used.
If the stock removal sensor is
installed it is not calibrated
correctly, please look in the manual
for the stock removal sensor.