Connect the plug on the additional dosing unit to the plug
located in the bottom of the dosing unit.
Remember the earth connections.
Replace the two cover plates.
Connect the white water hose to the ¾" thread.
Mount the straight end of the pressure hose onto the water
inlet tube on the back of AbraPol-10:
− Insert the filter gasket in the coupling nut with the flat
side against the pressure hose.
− Tighten the coupling nut completely.
Mount the other end of the pressure hose on the water mains
tap for cold water:
− Mount the reduction ring with gasket on the water mains
tap, if necessary.
− Introduce the gasket and tighten the coupling nut
completely.
Connect the silicone tubes to the inlets/outlets of the dosing
pumps.
Check that the reinforced hose does not interfere with the
belts.
Reconnect main power and allocate the bottles.
OP-X suspension can only be used in bottle 4.