Herunterladen Inhalt Inhalt Diese Seite drucken

Compatibility With Radiographic Generators - KaVo IXS Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für IXS:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 137
KaVo IXS
Environmental Conditions
USAGE
KaVo IXS sensor is not suitable
to be operated in oxygen rich
and/or explosive environments
TRANSPORTATION AND
STORAGE
Transport in the supplied
protective package
EU Classification
Protection against water/matter – IP 68

Compatibility with Radiographic Generators

KaVo IXS is generally compatible with any dental X-ray unit and generator capable
of supplying the required range of exposure times and doses.
Follow the instructions of the X-ray generator to set the desired dose range.
The radiation of X-ray tubes is controlled by the settings of:
Exposure time (msec) or pulses
Voltage (kV or kVp)
Current (mA)
Some controls allow for the modification of all of the above, some have fixed settings
for current and voltage.
Distance of X-ray Source to Sensor
There is a correlation between the distance of the X-ray source to the sensor and
the dose received by the KaVo IXS sensor. The radiation that reaches the sensor
decreases with the square of the distance. If you double this distance, you receive
only 1/4 of the radiation.
6-2
Humidity
Air
Pressure
30% to
700 to
95%
1060
hPa
10% to
95%
Class IIa medical device according to MDD
93/42/ECC
Ambient
Temperature
5° to 30°C
-40° to 70°C
English

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis