Herunterladen Inhalt Inhalt Diese Seite drucken

Chapter 2 - Safety And Disposal Procedures - KaVo IXS Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für IXS:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 137
Chapter
2
The device must be installed and used in accordance with the safety regulations and
instructions for use supplied in this User Manual, for the purposes and applications
for which it is intended.
All external surfaces of the sensor, sensor cable, sensor holders and the sensor
sheaths are considered to be applied parts and are safe for normal or accidental
patient contact during use.
The KaVo IXS has no serviceable parts and does not require calibration. Do not
open the device to service it. All aspects of the KaVo IXS that are meant to be
attended to by the operator are accessible without opening the internal components
of the device. If there is a service problem, contact a qualified dealer service
representative or KaVo Technical Support.
Important:
KaVo IXS is a USB compliant device and shall be used with USB compliant cables
suitable for high speed/USB 2.0 cables. Such cables are either marked "USB 2.0"
or "USB Hi-Speed."
USB certified hubs can be used to extend the distance to the USB host/computer.
The length of the cable connection to the hub or between hubs shall not exceed 5m.
Certified Hub (1.014.1145) is available for purchase and sold
NOTE:
separately.
CAUTION
Using non-USB compliant cables or hubs, or exceeding the maximum
count of USB hub devices for extending the distance, can degrade the
immunity of the KaVo IXS sensor to electromagnetic fields or increase
the emission of electromagnetic fields from the sensor.
English
Safety and Disposal
Procedures
Portable/mobile radio frequency communications equipment
can affect the function of the KaVo IXS sensor as well as any
other electronic medical equipment.
2-1

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis