Commissioning
PRACTICAL TIP
When processing solvents with a tendency toward boiling retar-
dation (e.g. ethanol), it is recommended to calibrate the sensor
slightly less sensitively.
This prevents the boiling delay from being detected as a fluid and
that the fluid level in the rotary piston decreases excessively before
adding new media.
The sensor should only react to liquids, never to boiling delay or
foam!
➜
Start the calibration process with
➜
The calibration process can be canceled at any time with
➜
The progress is visualized by the traffic light display. When the calibration is
complete, the green dot on the right lights up.
➜
Confirm the completion of the calibration process with
➜
After the sensor calibration has been completed, the start screen of the last active
operating mode is displayed.
.