Herunterladen Inhalt Inhalt Diese Seite drucken

Heidolph Hei-VOLUME Distimatic Pro Betriebsanleitung Seite 132

Vorschau ausblenden Andere Handbücher für Hei-VOLUME Distimatic Pro:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation
PRACTICAL TIP
Prerequisites
The following conditions must be met for rinsing the distillate tube:
Heating bath of the rotary evaporator is switched on
Recirculation chiller is switched on
Vacuum pump is switched on
Procedure
Depending on the selection, the various lines of the system are flushed.
To fill the evaporation flask, see also section "Commissioning" on page 96.
Select a vacuum value adequate for the solvent and set the
hysteresis to 50 mbar.
A large pressure range is passed through by the large
hysteresis.
All glass parts are first wetted, after which drops are formed,
which then run off.
As a result, the glass parts in the distillate area are cleaned
within a short time.
131